EXOGAM detector characterisation
The EXOGAM Clover detectors will be used with digital
electronics to allow the use of Pulse Shape Analysis (PSA) techniques to aid
the identification of the interaction position of incident gamma-ray
photons. The Liverpool group in collaboration with the EXOGAM collaboration
is leading the characterisation of the detector response function in order
to allow position information to be derived from the detector preamplifier
signals.
The Liverpool automated scanning table and digital
electronics system has been used to calibrate the response of the Ortec
EXOGAM Clover detectors utilising both singles and coincidence measurements.
The picture (above right) shows the intensity as a function of position for 662
keV photons incident on a Clover detector. The data was collected using a
2mm collimator and 5 minutes acquisition at each position. The image was
produced from the data by demanding all the energy is deposited inside a
outer contact segment. The effective segmentation of the detector is clearly
visible.
The rise time response of a detector is shown in the
figure (left) . The observed range of rise times is 100-220ns. The picture
shows the crystallographic lattice orientation of the germanium FFC
crystals. A specification of the EXOGAM detector is that the lattice
orientation should be aligned in all four crystals and that the 100 axis is
aligned for maximum rise time differences (as shown).
|